Published August 21, 2026 at 21:31
Most measuring instruments state their accuracy as a single figure. A stick micrometer set does something unusual: it states it as a formula. The accuracy here is ±(3 + n + L/50) µm, where n is the number of extension rods fitted and L is the measuring length in millimetres. That’s unusually candid, and it tells you more about how long internal measurements actually behave than an entire instruction manual would.
What it is
An internal stick micrometer set covering 50–1000 mm using eight extension rods that screw onto the micrometer body. The graduation is 0.01 mm, the measuring faces are spherical and made of carbide, the rod’s outer diameter is 15.5 mm, and it has a locking screw. A 50 mm setting standard is included, and everything is supplied in a wooden case.
The problem it solves
A stick micrometer is the simple solution to a difficult problem: measuring internal dimensions that are too large for almost anything else. A cylinder bore of 800 mm is beyond the reach of both a caliper and a three-point micrometer — but a rod extended to the right length gets there, and it can also be fed in through an opening far smaller than the dimension being measured.
The formula explains the price you pay for that reach. Three terms, three different sources of error:
- 3 µm is the base error — the micrometer body itself, its screw and measuring faces.
- + n µm is the cost per extension. Each joint is a threaded connection that must sit perfectly in line and completely free of play; in practice it adds one micrometre each. Measure 800 mm with seven extension rods fitted, and you’ve added seven micrometres from the number of joints alone.
- + L/50 µm is the length dependency — every fifty millimetres of measuring length costs one micrometre. This comes from the rod’s own deflection under its weight, thermal expansion, and the fact that a small angular misalignment produces a larger error the longer the instrument is.
Worked through a real case: a measurement of 1000 mm with eight extension rods comes out to roughly ±31 µm, while the same instrument at 100 mm with a single extension gives around ±6 µm. Same set, same screw, five times the uncertainty — purely because of the reach.
The takeaway is practical rather than discouraging: use as few extension rods as possible for the dimension you’re taking, and don’t expect hundredth-of-a-millimetre control on a metre-scale measurement. If the tolerance on a large bore needs to be tighter, a different method is needed — comparative measurement against a known master, or an instrument that isn’t built up from segments.
The fact that the measuring faces are spherical ties into the same issue. A sphere makes contact at a single point regardless of how the instrument is tilted, which lets you rock the rod and search for the largest reading — the true diameter — without the contact geometry changing during the search. With flat faces, a tilt would change both the contact point and the reading at the same time.
Three typical use cases
- Large cylinders and housings: internal diameters in pumps, compressors and hydraulic cylinders where other instruments fall short.
- Welded structures: checking shell diameters and openings in tanks and pipes, often through a small inspection hatch.
- Assembly and fitting: measuring the distance between surfaces inside a structure, where two parallel faces need to be checked against each other.
Why it pays off
A set covering 50–1000 mm replaces a whole row of instruments that would each only cover part of the range on their own — and costs less than several of them combined. For a workshop that measures large internal dimensions occasionally, but not daily, it’s usually the only economically sensible solution.
The second benefit is that you know what the measurement is worth. An instrument that states its own uncertainty as a formula lets you determine in advance whether the method is good enough for the tolerance on the drawing — instead of discovering afterwards that the reading could never have held up to what was assumed. That’s an unusual and underrated form of honesty in a datasheet.
Read more about the internal stick micrometer set 50–1000 mm →